KD Scientific Legato 200 Series Syringe Pump User’s Manual

Publication 5617-002-Rev F

Fuse Replacement

Key Information

1.

Make sure the power cord is disconnected

from the main supply before servicing the fuse.

2. Use only Type 3AG, 1/4" x 11/4, ½ amp,

250 volt; Type T (time delay) fuses (KD

Scientific Part No. 5146-037 or equivalent).

Turn off power and remove power cord from power

module. Use a straight blade screw-driver to pry open

the access door. Remove the fuse holder and then

remove the fuses from this holder as shown in figure

below. Replace fuses, and then replace the fuse holder.

1.

Remove screw and swing cover to the side.

2. Slide battery out from under clip.

3. Install Lithium Coin Battery: 3V, 16mm CR1620

(Part No. 5155-288 or equivalent) by sliding

under clip with positive side facing out

Battery Replacement

!

CAUTION: USE ONLY SAME TYPE &

RATED BATTERY. OBSERVE POLARITY

WHEN INSTALLING.

+