Function instructions – Klein Tools DTL304 User Manual

Page 7

Advertising
background image

7

• Press “ ” or “

” until the display shows “

”.

• Press “ENTER” to display “ ” or “

” in the time display.

If “

” is displayed, press “ ” or “

” to select “ ”.

• Press “ENTER” to display choice number 3 in the secondary

display and

00:30 h:m

in the time display.

• Press “ ” or “

” until the display shows the Auto-Power-

Off Time choice number you want, and then press “ENTER” to

store the Auto-Power-Off setting in memory.

• Auto-Power-Off mode is automatically disabled in MIN/MAX

and logging modes.

H.

Change Logging Interval

• The logging interval determines how often the thermometer

stores logged readings in memory. You choose the length of

the logging interval. The thermometer stores logged readings

at the end of each logging interval.

• Select logging intervals

• Press “ ” or “

” until the display shows

and “ LOG ”.

• Press “ENTER” to display the logging interval choices.

• Press “ ” or “

” until the display shows the logging

interval you want, and then press “ENTER” to select.

• For user-defined logging interval (1 sec. to 23 hrs : 59 mins).

Press “ ” or “

” until the display shows “h:m” or “m:s”,

and then press “ENTER” to select. The last number in the

time

display

blinks.

Press “ ” or “

” until the logging interval you want

appears on the display, and then press “ENTER” to select.

• Holding down “ ” or “

” causes the number to change

more quickly.

I.

Setting Temperature Offset

• You can adjust the thermometer’s readings to compensate

for the errors of a specific thermocouple. The allowable

compensation range is ±5.0°C or K, and ±9.0°F. The DTL304

allows you to store individual offsets for T1, T2, T3 and T4.

• Press “ ” or “

” until the display shows “ OFFSET ”.

• Press “ENTER” to display T1 reading (blinking), T1 offset.

• Press “ ” or “

” to display T1, T2, T3 and T4 in the primary

display to select 1 input.

• Press “ ” or “

” until the primary display shows the

correct reading, and then press “ENTER” to store the offset

setting in memory.

• NOTE: Reset the offset to 0.0 when it is no longer needed.

• NOTE: If T1, T2, T3 or T4 has no offset, “ OFFSET ”will not be

displayed.

J.

Resetting The Thermometer

• To restore to factory settings, press “ ” or “

” until

the display shows “

” then press “ENTER” to display

“ in the primary display. Press “EXIT” to exit setup.

FUNCTION INSTRUCTIONS

1.

Connecting Thermocouples

• Thermocouples are color coded based on the ANSI color code:

• Plug a thermocouple into the input terminal(s).

• Set the thermometer for the correct thermocouple type. To

change the thermocouple type, see:

Set Up Instructions section,

Set Up Instructions

Set Up Instructions

F: Thermocouple Type

1.

:01 sec

2.

:10 sec

3.

1:00 min

4.

10:00 min

User Defined

J Type = Black
K Type = Yellow

T Type = Blue
E Type = Purple