2.

Displaying Temperature

• Press “HOLD (°C/°F/K)” for more than 1 second to select the

correct temperature unit.

• Hold or attach the thermocouple(s) to the measurement

location. The temperature reading appears in the selected

display.

• Press “T1/T2 T3/T4 (T1-T2, T3-T4)” momentarily to toggle T1/

T2 or T3/T4 reading in the primary display.

• Press “T1/T2 T3/T4 (T1-T2, T3-T4)” for more than 1 second

to scroll T1-T2, T3-T4 or the internal temperature of the

thermometer in the secondary display.

• NOTE: The display shows “----” when a thermocouple is not

NOTE:

NOTE:

connected. The display shows “

” (OVERLOAD) when the

temperature being measured is outside the thermocouple’s

valid range.

3.

Holding the Displayed Temperature

• Press “HOLD (°C/°F/K)” momentarily to freeze the readings on

the display. The display shows “ HOLD ”.

• Press “HOLD (°C/°F/K)” momentarily again to turn off the “

HOLD ” function.

4.

Viewing the MIN/MAX/AVG Readings

• NOTE: MIN/MAX and AVG values are captured only while this

NOTE:

NOTE:

mode is active. Exiting the MIN/MAX/AVG mode will clear all

values.

• Press “MAX/MIN” momentarily to step through the MIN

(minimum),MAX (maximum), or the AVG (average) readings in

the secondary display.

• The elapsed time since entering MAX/MIN mode appears with

the average reading.

• The time at which the MINIMUM or MAXIMUM occurred

appears with the MINIMUM or MAXIMUM reading on the

display.

• Press “T1/T2 T3/T4 (T1-T2, T3-T4)” momentarily to toggle

showing the maximum (MAX) of T1, T2, T3 or T4 readings and

its displayed time.

• Press “T1/T2 T3/T4 (T1-T2, T3-T4)” for more than 1 second to

show the maximum (MAX) of T1-T2 or T3-T4 reading and its

displayed time on the secondary display.

• The minimum (MIN) or average (AVG) reading can be

displayed in the similar way.

• Press “MAX/MIN” for more than 1 second or “

” button

for more than 1 second to exit MIN/MAX mode.

5.

Using the Offset to Compensate for Probe Errors

• Use the offset option in “SETUP” to adjust temperature

readings to compensate for the errors of a specified

thermocouple.

• Plug the thermocouple into the input terminal.

• Place the thermocouple in a known stable temperature

environment (such as an ice bath or a dry well calibrator).

• Allow the readings to stabilize.

• In “SETUP”, change the offset until the primary display

reading matches the calibration temperature, see:

Set Up Instructions section,

Set Up Instructions

Set Up Instructions

I : Setting Temperature Offset

6.

Using Memory

• During logging session, the thermometer stores logged

readings in its memory. At the end of the logging session you

can view the logged readings on the display. You can also

transfer the logged readings to a PC running the optional

software, which displays the readings on an online form,

which can be printed or stored for later use.