Labnet Z400K High Capacity Centrifuge User Manual

Page 15

Advertising
background image

Z 400 K / March 1997

15 E

4.8

Keyboard - Starting the centrifuge - "quick"-key

1

Key "lid": to open the centrifuge lid. If the green LED

2

of the key "lid" is lighting the lid is closed correctly.

3

Key "quick": for short time runs

4

Key "start": to start the centrifuge.

5

Key "stop": to stop the centrifuge before the preset
operating time has expired or to stop the centrifuge
at continuous run.

Figure 6

Starting the centrifuge

Starting the centrifuge

Starting the centrifuge

Starting the centrifuge

Starting the centrifuge

The rotor has to be fixed correctly and completely loaded (see point 3).
Close the centrifuge lid. As soon as the green LED of the key "lid" is lighting the
centrifuge can be started. Therefore press key "start".

"quick"-key - Short time runs
For short centrifuge runs you can start the run with the key "quick". Press the
key "quick". The centrifuge starts and keeps running as long as you press the
"quick"-key. The operating time will be indicated in seconds on the digital
indication "time".

4.9

Key "stop"

Press the key "stop" if you want to interrupt a centrifuge run. The centrifuge
decelerates according to the adjusted brake intensity.

Advertising