Labnet Z400K High Capacity Centrifuge User Manual

Page 20

Advertising
background image

20 E

Z 400 K / March 1997

Error no.: 11

Error no.: 11

Error no.: 11

Error no.: 11

Error no.: 11

Temperature sensor

Temperature sensor

Temperature sensor

Temperature sensor

Temperature sensor

Reason:

Short circuit at the temperature sensor or at the sensor cable

Action:

Call service

Reason:

Chamber temperature is too low, below - 25°C

Action:

Solenoid valve is not working. Call service

Error no.: 20

Error no.: 20

Error no.: 20

Error no.: 20

Error no.: 20

No rotor identification

No rotor identification

No rotor identification

No rotor identification

No rotor identification

Reason:

No rotor inserted

Action:

Insert rotor into the unit

Reason:

Rotor identification sensor defective

Action:

Call service

Reason:

Inserted rotor has no indicator ring

Action:

Use a correct rotor

Reason:

Rotor is not fixed correctly to the motor shaft

Action:

Insert rotor correctly. The pins have to align correctly with the

rotor slots (see chapter 3)

Error no.: 21

Error no.: 21

Error no.: 21

Error no.: 21

Error no.: 21

Start Bit is missing

Start Bit is missing

Start Bit is missing

Start Bit is missing

Start Bit is missing

Reason:

A magnet of the indicator ring is missing

Action:

Check indicator ring and call service

Error no.:

Error no.:

Error no.:

Error no.:

Error no.: 10

10

10

10

10

Overtemperature in the rotor chamber (more than + 50 ° C)

Overtemperature in the rotor chamber (more than + 50 ° C)

Overtemperature in the rotor chamber (more than + 50 ° C)

Overtemperature in the rotor chamber (more than + 50 ° C)

Overtemperature in the rotor chamber (more than + 50 ° C)

Reason:

Breakdown of the refrigeration system / Electronic failure

Action:

Call service

Reason:

Temperature sensor defective / Electronic failure

Action:

Call service

Advertising