33

Leica CM3050S – Cryostat

5. Operation

These parameters can be displayed via the ar-

row keys.

When pressing

, the parameters are dis-

played in the sequence listed on this page;

when pressing

, they are displayed in re-

verse order. The display can be scrolled contin-

uously, i.e. after the last parameter the display

will automatically start again with the first.

1. Refrigeration state

Meaning of the refrigeration symbols (asterisks):

--> see chapter 5.3.5 ‘Status indication’.

2. Section thickness

The sectioning thickness is set with th

o

key (see chapter 5.4.7 ‘Setting the section thick-

ness’).

3.

Section counter

Once displayed (via arrow keys), the section

counter can be set to ‘0’ at any time.
• To set to ‘0’ pres

.

If the section counter is not set to ‘0’ by the

user at some point, sections are counted up to

‘999’ at which point the counter automatically

switches back to ‘0’.

5.3.6 Indication of refrigeration state,

section thickness setting

and counters

4.

Total section thickness counter

Can also be set to ‘0’ at any time (via

key).

If the total section thickness indication is not

set to ‘0’ manually, the instrument counts up to

maximum 9999.5 µm at which point the counter

will automatically switch back to ‘0’.

This indication shows the total thickness of all

sections carried out since the counter was last

set to ‘0’. (All sections, in sectioning and trim-

ming mode, are counted).

5.

Preset counter

Setting preset counter to ‘0’:
Press

key.

can also be pressed while the continuous

sectioning mode is in motion; in this case the

preset counter is set to ‘0’ and sectioning stops

upon reaching the lower limit position.

After it has been set to ‘0’, the preset counter is

inactive.

Reloading the former reference value:
Press

and

simultaneously.

Reloading the former reference value is possi-

ble also while the continuous sectioning mode

is in motion:

Upon pressing both arrow keys, the instrument

continues sectioning without interruption, start-

ing to count down from the old reference value.