5 coarse feed keys, Operation – Leica Biosystems CM3050 S User Manual

Page 38

Advertising
background image

38

Instructions for Use V 1.7 RevC - 08/2014

5. Operation

5.4.5 Coarse feed keys

The four coarse feed keys in functional area B

function as follows:

Coarse feed ‘fast backward’

- Travelling speed 1,000 µm/s

• Press

key to activate.

Specimen head travels backward

(= away from knife):

a. until it reaches the rear limit position,
b. until it is stopped by pressing one of the

coarse feed keys (to stop the function,

any of the four coarse feed keys can be

pressed).

Coarse feed ‘slow backward’

- Travelling speed 500 µm/s

• To start, press

and hold.

Specimen head travels backward (= away

from knife) as long as the key is held.

As long as the backward coarse feed (both fast

and slow) is in motion, the ‘HOME’ LED is illumi-

nated.
Upon reaching the rear limit position, the

‘HOME’ LED starts flashing and the display in

control panel 1 reads ‘HOME’.

Coarse feed ‘fast forward’

- Travelling speed 1,000 µm/s

• To activate the function, press

and hold.

The specimen head travels forward (= to-

wards the knife) as long as the key is held.

Coarse feed ‘slow forward’

- Travelling speed 500 µm/s

• To activate the function, press

and hold.

The specimen head travels forward (= to-

wards the knife) as long as the key is held.

As long as the forward coarse feed (both fast

and slow) is in motion, the ‘STOP’ LED is illumi-

nated.
Upon reaching the front limit position, the

‘STOP’ LED starts flashing and the display in

control panel 1 reads ‘STOP’.

LED ‘HOME’ (located in key

‘Coarse feed fast backward’)

‘STOP’-LED (located in key

‘Coarse feed fast forward’)

Functional area B - all models

Advertising