4 available parameters – Memmert AtmoCONTROL User Manual

Page 16

Advertising
background image

16

D24042 | Date 09/2014

AtmoCONTROL

4.1.4 Available parameters

Below, all the parameter icons with their adjustment options are shown.

Which parameters are available to adjust the programme depends on the appliance for

which a programme is to be created. Only those parameters are available that the appli-

ance is able to implement. For appliances without humidity regulation, for example, no

humidity icon is available. The respective adjustment options (temperature ranges etc.) are

appliance-specific.

Broad parameter representation

parameter representation

Depiction

in icon bar

Meaning

Depiction on

editor thread

Function and

adjustment options

?

15

STAND BY

15

Hold tem-

perature

Tolerance band

Function

Maintains a set temperature for a

specific time.

Adjustment options

Adjustment options

Name of programme segment

1

Duration (time or infinite ∞)

Temperature to be maintained

Tolerance value above/below

Alarm if limits are exceeded

Safe

2

(For programme example, see page

24)

?

15

STAND BY

15

Change

tempera-

ture

0,1 K/min

Tolerance band

Function

Increases or decreases temperature

over a specific time to a set value.

Adjustment options

Adjustment options

Name of programme segment

1

Duration

Target (setpoint) temperature

Tolerance value above/below

SPWT

3

(For programme example, see page

25)

1

When run, this is displayed in the status bar of the appliance

2

When Safe is "on", it is ensured that the value really is maintained within the tolerance band as long as

specified, and only then is the programme continued (this is sensible for sterilisers, for example). If the

actual values leaves the tolerance band, the clock timer starts again from the beginning.

3

SPWT: Setpoint wait. If this is "on", the programme sequence is not continued before the setpoint

value is reached, even if the set time has already expired. If this is "off", the programme sequence is

continued after the set time has expired, irrespective of whether the setpoint value was reached or not.