2 programme example with door locking – Memmert AtmoCONTROL User Manual

Page 25

Advertising
background image

D24042 | Date 09/2014

25

AtmoCONTROL

4.4.2 Programme example with door locking

1

2

5

8

6

3

4

Tolerance band

1,7 K/min

7

95 °C

20 °C

45 °C

12:00 h ≥ 0:30 h

1

2

3

5

6

Fig. 9

The door is locked at the beginning of the programme (1). Then, the appliance heats up to

Fig. 9

Fig. 9

95.0 °C (2) and maintains this temperature for 12 hours (3). Subsequently, the temperature is

The door is locked at the beginning of the programme (1). Then, the appliance heats up to

The door is locked at the beginning of the programme (1). Then, the appliance heats up to

lowered (5) for 30 minutes (4) to 45.0

C (2) and maintains this temperature for 12 hours (3). Subsequently, the temperature is

C (2) and maintains this temperature for 12 hours (3). Subsequently, the temperature is

°C and then, the door is opened (6). The setting "

C (2) and maintains this temperature for 12 hours (3). Subsequently, the temperature is

C (2) and maintains this temperature for 12 hours (3). Subsequently, the temperature is

SPWT

on" (8) ensures that the door is opened only when the temperature really has dropped to

C and then, the door is opened (6). The setting "

C and then, the door is opened (6). The setting "

45.0 °C, even if this takes longer than 30 minutes. Below the temperature change is shown in

on" (8) ensures that the door is opened only when the temperature really has dropped to

on" (8) ensures that the door is opened only when the temperature really has dropped to

K/min (7).

C, even if this takes longer than 30 minutes. Below the temperature change is shown in

C, even if this takes longer than 30 minutes. Below the temperature change is shown in