8. Event codes of the log fi le Log.txt

(see page 12)

Error codes for Generation 2012 appliances

Status: October 09, 2012

In case of a hardware error of the oven, the controller displays the following error codes in a status /
error window, as well as logs them in the “Log.txt” file in the “Config” directory on the SD card.

The position of the error number further specifies the position of the error.

Error code

Parameter

Description

101: OS Error

Operating system error

102: File System Error

File system error

103: USB Error

USB error

104: GUI Error

Graphical user interface error

105: IP Stack Error

Ethernet error

106: I2C Bus Error

12C bus error

107: RTC Error

Realtime clock error

108: RAM Disk Error

RAM disk error

109: WatchDog Reset

Watchdog error

Power Supply OK

110: Power Supply OK

111: Controller Restart

Restart after reset

201: Config Error

Configuration error

202: Calib User Error

Error in calibration file (user)

203: Calib System Error

Error in calibration file (calibration system)

204: PID Config Error

Error in PID parameters

205: User Config Error

Error in user settings

206: Battery Error

Battery low

207: SD Card Space

Warning at 95% disk space usage

208: SD Card Full

Error, SD card is full

209: SD Card Missing

SD card is missing

210: Failed To Copy Protocol

Error copying the protocol

211: Restauration Failed

Error restoring LastState

212: Max Count Of Profiles

Maximum number of profiles on SD card reached

301: Fan Error

Fan speed error

302: Heating Error

Heating error

303: Temp Limitor

Temperature limiter has triggered

304: Door open

Door is opened

305: Hzg Err

2

00000

Triac optocoupler heating 1 power module 1

defective

0

2

0000

Triac optocoupler heating 2 power module 1

defective

Note:

The position of the red digit

indicates the defective stage

of the appliance.

00

2

000

Triac optocoupler heating 1 power module 2

defective

000

2

00

Triac optocoupler heating 2 power module 2

defective

0000

2

0

Triac optocoupler heating 1 power module 3

defective

00000

2

Triac optocoupler heating 2 power module 3

defective

1

00000

Triac heating 1 power module 1 defective

0

1

0000

Triac heating 2 power module 1 defective

00

1

000

Triac heating 1 power module 2 defective

000

1

00

Triac heating 2 power module 2 defective