Especially with heavy loading of the steriliser it is not sufficient to use these typical values without further
tests. Reliable sterilisation requires validation of the individual sterilisation process, e.g. with the aid of
additional temperature probes or by using biological or chemical indicators.

Note:
In sterilisation processes the vent valve on the oven must be closed after the moist
sterilisation load has been dried!

10.4 Steriliser cassettes

The cassettes should preferably be so arranged in the steriliser that the hot air flow can pass readily through
the air slots.

The load to be sterilised is placed into the steriliser cassettes wrapped in aluminium foil or in steriliser foil
suitable for hot air

(as in the Table in the Section „Guidelines for sterilising“)

. The air slots in the cassette

must be open during sterilisation.
A temperature probe to confirm the temperature of the load can be introduced through the opening (2).
After sterilisation has been completed the air slots must be closed by moving the slide knob (1).
The sterilised and packed load can then be stored briefly in the closed cassette.