Peak/Valley Change Detector Process

MTS MultiPurpose TestWare®

Event Process Descriptions

371

For more information

For more information about triggers, see

“About Sequencing Processes”

on page

87.

How to Define a Peak/Valley Change Detector Process

1. Drag the

icon from the Process Palette to your procedure table.

2. Click the process icon in your procedure table or in the navigation pane.

3. Click the Settings tab.

4. In the Signal list, click the signal you want to monitor.

5. In the Sensitivity box, set the units, and then type a value.

6. In the Tolerance box, set the units, and then type a value.

7. Click the Options tab.

8. Optional—If you want to delay the Peak/Valley monitoring process for a

number of cycles to allow for a “settling in” period, type a value in the
Enable After box.

9. Under Reference Values, select Default or User-specified.

If you select User-specified Reference Values, type the desired Upper and
Lower Reference values.

10. Under Trigger Mode, set the trigger mode.

11. Optional—Enable (check) the Log Message As control and select the type

of message you wish to generate.

Counter Type

Enables a process counter.

Options are None, Transient, or Fixed. Transient counters are displayed only
when the process is active. Fixed counters are displayed throughout the test.

Counter Label

Names the counter.

When you run the test, this name appears under Sequence Counters on the

MPT control panel.

Note

If this box is blank, the process name is used as the counter label.