Pump reset, Quick start instructions, Operating the pump – New Era NE-500 Multi-Phaser Programmable OEM Syringe User Manual

Page 2

Advertising
background image

New Era Pump Systems Inc.

www.SyringePump.com

Model NE-500/NE-501 Multi-Phaser™

Publication #1200-02

i

09/18/14

Quick Start Instructions

Assumes that the pump was not previous programmed with a multiple Phase Pumping Program.

Plug in the pump.

Attach the computer cable to the "Computer" connection on the RJ-11 connector.

Start the terminal emulator program on the computer (downloadable from the web site).

Baud Rate: 19,200

Data frame: 8N1

(Default settings for the downloadable terminal emulators)

Note: Each command sent to the pump must be terminated by a carriage return (0x0D) (Sent automatically by
terminal emulator). Space characters are ignored. Characters can be upper or /lower case.

Note: Wait for the response from each command before sending the next command.

Setup Pumping Parameters

Set the Syringe Inside Diameter:

Where "nn.nn" is the syringe inside diameter in mm: Send the command: DIA nn.nn

Set the Pumping Rate.

Where "nn.nn" is the pumping rate and "xx" is the rate's units: Send the command: RAT nn.nn xx

Rate units "xx" value

mL/hr

MH

µL/hr

UH

mL/min

MM

µL/min

UM

If the pumping rate is out of range the response message to the computer will include "?OOR"

Set the Volume to be Dispensed or Continuous Pumping

Where "nn.nn" is the volume to be dispensed, send the command: VOL nn.nn

If the volume to be dispensed is 0, then the pump will pump continuously.

Note: The default units for the volume to be dispensed will be mL or µL, depending on the syringe size.
At or below 14.00 mm syringe diameter, volume units are µL.
Above 14.00 mm syringe diameter, volume units are mL. The default volume units can be overridden.

Set the Pumping Direction

For infusion, send the command:

DIR INF

For Withdrawal, send the command:

DIR WDR

Load the Syringe

Press in the white drive-nut button to move the pusher block.

Insert the syringe plunger in the pusher block slot.

Insert the syringe barrel flange in the flange brackets with the syringe barrel holder on the syringe. Tighten the
flange brackets onto the syringe flange. Tighten the pusher block screw.

Operating the Pump

Start the Pump:

Send the command: RUN

Stop the Pump:

Send the command: STP

Query the Pump Status

Send just a carriage return to the pump. The pump will respond with a status prompt

When Pumping

The pumping rate can be changed.

With continuous pumping, the pumping direction can be changed.

PUMP RESET:

PUMP RESET:

PUMP RESET:

PUMP RESET:

Send the command: *RESET