New Era NE-500 Multi-Phaser Programmable OEM Syringe User Manual

Page 40

Advertising
background image

New Era Pump Systems Inc.

www.SyringePump.com

Model NE-500/NE-501 Multi-Phaser™

Publication #1200-02

35

09/18/14

8.5.4

General Control and Status Commands

TTL I/O OUTPUT SETTING
OUT <n> <TTL level>
E { 1 |2 | 3 | 4 | 5 } <TTL level>

Expansion Port Connector

Sets TTL level on user definable output pin on the ‘TTL I/O’ connector.

<n> Indicates pin number on ‘TTL I/O’ connector

Valid value: 5 (Program Output pin)

E Sets pin on Expansion Port I/O connector on PCB.

Query is not applicable.

TTL INPUT QUERY
IN <n>
E { 1 |2 | 3 | 4 | 5 }

Expansion Port Connector

Queries TTL level of pin on ‘TTL I/O’ connector. Set is not applicable.

<n> Indicates pin number on ‘TTL I/O’ connector

Valid values: 2, 3, 4, and 6.

E Reads from the Expansion Port I/O connector on the PCB.

Query response: <TTL level>

BUZZER
BUZ [ 0 | { 1 [ < n > ] } ]

Sets / queries buzzer

Set:

0 = Turn buzzer off;

1 = Turn buzzer on

if <n> specified

If <n> = 0, buzzer beeps continuously,

otherwise buzzer beeps <n> times

if <n> not specified, buzzer sounds continuously

Query response: { 0 | 1 }

0 = Buzzer off
1 = Buzzer is on continuously or beeping.

8.5.5

System Commands

SET PUMP NETWORK ADDRESS AND BAUD RATE
* ADR [ <address> [ B { 19200 | 9600 | 2400 | 1200 | 300 } ] ]
[DUAL | RECP | ALTR]

Set/query pump network address

<address> => <n> [<n>]
<address> Valid range: 0 to 99

B { 19200 | 9600 | 2400 | 1200 | 300 } will change the baud rate as indicated. NOTE: The command

response and all further communications will be at the specified baud rate.

Special communications modes

DUAL Sets the pump to Dual Pumping mode with a secondary pump.

RECP Sets the pump to Reciprocating Pumping mode with a secondary pump.

ALTR Sets Alternating pumping mode.

All special communications modes use 19,200 baud rate.