Calibration – NTi Audio XL2 Sound Level Meter User Manual

Page 137

Advertising
background image

137

Calibration

5

Level Type

LZF

Default setting.

LHP100F Suppressing any low frequency noise, for

example wind during outdoor calibrations.

6

Electronic Data Sheet of Measurement Microphone

Read out data from the connected microphone.

User Calibration - With External Precision Calibrator

User-sensitivity setting for measurement microphones with an
external calibrator.

Turn the rotary wheel

to select the parameter Calibration

Level

4

and press enter .

Set the calibration level according to the calibrator used.

Generate the specific reference signal with the calibrator

onto the microphone.
Turn the rotary wheel

to select the parameter RUN next

to User Calibration

3

and press enter .

The pop-up window

Calibration: Calibration running ...

shows up and changes to Calibration: Successfully fin-
ished!
for a successfully completed calibration.

The user calibration is completed.

User Calibration - Manual Sensitivity Setting

In case no external calibration device is available, set the sensi-
tivity of non-NTi Audio microphones manually according to the
product specifications as follows:

Select

Sensitivity

2

with the rotary wheel .

Press enter

and adjust the sensitivity with the rotary

wheel in 0.1 mV/Pa steps. The setting range 100 µV/Pa -
9.99 V/Pa is supported.
Confirm with enter

.

The manual user calibration is completed.

User Sensitivity

The XL2 writes the user-sensitivity setting onto
the electronic data sheet of the connected NTi
Audio measurement microphone or microphone
preamplifier. All subsequent measurements au-
tomatically utilize this user-sensitivity.