Capture + Tolerances

Advanced Start Functions

You may start the passed/failed measurement by one of the
following possibilities:

Press the XL2 start button

.

Activate the automated level trigger in the tolerance file.

Activate the digital input 1 via the external I/O interface; e.g.

a foot switch or a PLC control.

Digital I/O Interface

Result Passed:

Output 1

Result Failed:

Output 3

Start Measurement:

Input 1

Tolerance Management

Tolerance curves can either be imported from txt-files or direct-
ly derived from captured measurements. The XL2 calculates
tolerance bands based on

Single captured results

Manually generated txt-files on the PC

Mean average of multiple captures

Min/Max curves of multiple captures

Create Tolerance Files Manually on PC

Capture a spectrum on the XL2 Analyzer.

Enter the tolerance mode and generate a tolerance band for

the capture.
Select the Capture & Start Tolerance Mode symbol.

Confirm with enter

.

Select

Save tol. to SD Card to export the capture tolerance

file. This file includes the reference capture and the tolerance
data. The XL2 generates a txt-file in the folder “Tolerances”.
Open the capture tolerance file with a text editor or MS Ex-

cel.
Edit the tolerance data according your requirements. You

may delete any unused frequencies. All remaining frequency
data has to remain the same. Any frequency changes will be
discarded by the XL2.