16

17

Input Filters

Depending on the selected measurement function the following sets of

filters are available:

SPL/RTA function:

Filters

Applications

FLAT

Flat frequency response

(no filtering)

Overall sound pressure level; all

sound signal components are included

without weighting - required for special

applications.

A-WTD

A-weighting filter in

accordance with IEC 61672

Applicable to most common sound

pressure levels if level weighting

compensation for human loudness

perception is required.

C-WTD

C-weighting filter in

accordance with IEC 61672

Applicable to higher sound levels if level

weighting compensation for human

loudness perception is required.

X-CRV

-1

Inverted X-Curve filter in

accordance with ISO 2969

Required for cinema installations

RLB (Revised Low-

frequency B-curve)

Broadcast Loudness

filter in accordance with

AES Paper “Loudness

Assessment of Music and

Speech” 2004

Applicable for loudness measurement

of broadcast material. The revised

B-weighted Leq measurement (RLB)

has the best correlation to the subjective

experience of humans.