NTi Audio Acoustilyzer AL1 User Manual

Page 26

Advertising
background image

26

27

Pause: The Leq, LCpeak and MAX/MIN measurements can be paused
• Select the PAUSE symbol and press the enter key.

• The PAUSE symbol flashes.
• The actual SPL value is continuously measured and displayed.
• Select the pause symbol and press the enter key to continue the

measurement; the clock continues to count.

Reset: Pressing the RESET icon restarts the measurement. The timer and

test results are reset.

Test Cycle Period: Applicable in SINGLE and REPEAT timer status only.

• SINGLE: Set to 10 min by default
• REPEAT: Set to 10 sec by default

Measurement Timer: Actual measurement in hours:minutes:seconds.

Big and Small Result: Two of the previously-listed measurements may be

display at the same time.

Time Weighting: All SPL measurements (SPL and MAX/MIN SPL)

are time-weighted measurements. Changes in the SPL level are

smoothed, resulting in quicker or slower change of the displayed value.

Available time weightings, corresponding to IEC 61672, are

• SLOW (long attack and release response time, t = 1s)
• FAST (short attack and release response time, t = 125ms)

The time weighting for the broad band values (

) and the RTA (

)

are always the same.

Overload

The OVL (Overload) flashes if the input

voltage exceeds the selected input range

of the instrument. As long as the OVL is

flashing the displayed Leq and LCpeak

results are not accurate.