NTi Audio Acoustilyzer AL1 User Manual

Page 17

Advertising
background image

16

17

RMS/THD function:

Filters

Applications

FLAT

Flat frequency response

(no filtering)

Overall sound pressure level; all

sound signal components are included

without weighting - required for special

applications.

A-WTD

A-weighting filter in

accordance with IEC 61672

Applicable to most common sound

pressure levels if level weighting

compensation for human loudness

perception is required. For example,

measuring residual noise of a device in

sound broadcasting applications.

C-WTD

C-weighting filter in

accordance with IEC 61672

Applicable to higher sound levels if level

weighting compensation for human

loudness perception is required - for

special applications.

HP400

Highpass 400 Hz in

accordance with DIN 45045,

-120dB/dec.

Removes any mains frequency (50/60Hz)

components from the test signal.

HP19k

Highpass 19 kHz

Removes any low frequency components

from the test signal, e.g. to measure

critical public announcements systems

with no audible change to the signal

(using a 20kHz pilot tone level)

Advertising