Nleq – NTi Audio Acoustilyzer AL1 User Manual
Page 33
32
33
⎟⎟
⎠
⎞
⎜⎜
⎝
⎛
⋅
⋅
=
∑
=
N
n
dt
LEQ
n
N
LEQ
1
10
_
10
1
log
10
MIN_dt Minimum SPL value measured within the actual log interval
∆t;
time weighted value (SLOW/FAST).
MAX_dt Maximum SPL value measured within the actual log interval
∆t;
time weighted value (SLOW/FAST).
Example of an SPL/LEQ log visualized with Microsoft Excel
(without LCpeak)
LCpeak Maximum C-weighted peak sound pressure level during the
whole measurement period. The measurement period is
indicated and controlled by the measurement timer.
LCpeak_dt Maximum C-weighted peak sound pressure level during the
log interval
∆t
.
SPL_Act Indicates the actual SPL level measured exactly at the end
of the log interval. Note that Leq_dt, MIN_dt and MAX_dt
characterize an interval more appropriately than this single
SPL_Act value.