Model 530 Operating Instructions

Quick Reference Bar Graph cont.

HART

®

Compatibility Mode -

ON/OFF (default)


The Model 530 has a HART

®

compatibility mode. This mode is useful when the devices being powered

communicate using the HART

®

protocol. In this mode the Model 530 connects a 250 Ω load resistor in

series with the output in both Source and Power Measure 2-Wire transmitter modes. This eliminates the
requirement of an external 250 Ω load resistor. This resistor is typically shown in connection diagrams and

manuals for HART

®

devices.

If HART

®

Compatibility Mode is ON, a 250 Ω load resistor is automatically switched in series with the

output in Source and Power Measure 2-Wire Transmitter modes. The output compliance with HART

®

Compatibility Mode ON is 950 Ω at 20 mA.

If HART

®

Compatibility Mode is OFF there is no 250 Ω load resistor in series with the output. This will

increase the output compliance voltage to drive 1200 Ω at 20 mA.

EZ-Check
HI/LO Readings ON/OFF

(default)


If the EZ-Check HI/LO Readings option is ON, the highest and lowest readings will automatically be

saved in the HI and LO EZ-Check™ positions.

If this option is OFF the HI and LO positions will show the current reading.

Factory Reset

ON/OFF

(default)

If Factory Reset is ON, the unit will restore all factory defaults when the Model 530 is turned OFF and back
ON. This will reset any changes made in the Model 530 Configuration options, returning the unit to its

simplest factory configuration.

Instructions for Enabling and Disabling the Configuration options in the Model 530

1. Turn the Model 530 on.
2. Press the EZ-Dial knob while the “PRESS EZ-DIAL KNOB FOR CONFIGURATION” message is

displayed.

3. Select options by turning the EZ-Dial knob until the arrow points to the desired option.
4. The option can be enabled or disabled by tapping the EZ-Dial knob.

5. Turning the Model 530 off to exit configuration.

Error

Conditions

Source Milliamps:

“HIGH

Ω” flashes in place of bar graph

Power Measure:

“CURRENT LIMITED” flashes in place of bar graph

2-Wire Transmitter Simulate:

“LOW SUPPLY” flashes in place of bar graph


Read Milliamps:

“CURRENT LIMITED” flashes in place of bar graph

Read Volts:

“OVERRANGE” flashes in place of bar graph