Model 530 operating instructions, Source mode, Wire transmitter simulation mode – Practical Instrument Electronics 530 User Manual

Page 5

Advertising
background image

Model 530 Operating Instructions

Source

Mode

Transmitter Input

(Sensor)

(Process Signal)

(Simulated Input)

Receiver

Power Supply

0 to 60 VDC

Typical

2-Wire

Transmitter

+ IN -

REF +OUT-

Source mode uses internal power to supply current from 0.000-24.000 mA into as much as 1200Ω (with
HART resistor disabled) until the end of battery life. The calibrator will indicate “HIGH Ω” if connected
improperly. The three-position EZ-Check switch provides instant zero and span calibration outputs. EZ-

Checks from 0-24 mA can be stored in the zero and span switch positions. The output is adjusted in 0.001
or 0.100 ma increments (0.01 or 1.00% in percent display units) with the EZ-Dial knob. Step the output

by pressing the EZ-Step button. Six different user-selected steps sizes are available. Hold the EZ-Step
button to activate hands-free auto step/ramp feature.

2-Wire Transmitter Simulation Mode

Transmitter Input

(Sensor)

(Process Signal)

(Simulated Input)

Receiver

Power Supply

0 to 60 VDC

Typical

2-Wire

Transmitter

+ IN -

REF +OUT-

The Model 530 can simulate a 2-wire transmitter in the 4-20 mA process loop. The calibrator will indicate
“LOW SUPPLY” if improperly connected. The EZ-Check switch and EZ-Dial knob allow rapid and fine control

of loop current. The EZ-Step button and hands-free auto/ramp function allow a complete check of
calibration points.


841 Holt Road #1 Webster, NY 14580

Tel: 585-872-9350 Fax: 585-872-2638

[email protected] http://www.piecal.com/

Practical Instrument Electronics, Inc. Copyright 2002. All rights reserved.

530-9002 2/26/04

5-8