0003

General points

Cleaning

Do not clean the ZM200 with flowing water.

Mortal danger from electric shock

Only use a cloth dampened with water.

Solvents are not permitted.

Maintenance

1

2

3

4

5

6

The following service work should be performed from time
to time, although at the latest once per month, in order to
guarantee the operational reliability of your ZM200:

Check roller 1 of the trannion piece for easy
movement and oil, if necessary, e.g. with sewing
machine oil

Clean magnets 2 on the trannion piece

Clean pressure pieces 3 in the cover

Clean motor shaft 4 under the labyrinth disc but do
not use compressed air. We recommend using a
suitable brush.

Replace air filter 5 at the rear.

Remove any dirt that has accumulated in the space
under the perforated sheet 6 on the bottom of the
device, unscrew perforated sheet 6 for this purpose

The secure closing of the ZM200 housing cover depends on
the easy movement of roller 1 on the trannion piece.

These operating instructions do not contain any repair
instructions. In the interests of your own safety, repairs
should only be performed by Retsch GmbH, an authorised
representative or by Retsch service technicians.