Adjusting the lock if the cassette
cover leaks

1

2

3

4

5

The lock in the housing cover can be subsequently adjusted to rectify leaks in the cassette cover.

Apply open-jaw spanner 1 size 30, not contained in the scope of supply

Undo lock nut 2 by turning it to the left

Turn counter nut 3 by hand in ¼ turns so that the distance between lock roller 4 and lock nut 2
becomes smaller

Repeat the steps above until the cassette cover is sealed again

CAUTION !

Once the adjustment work has been completed, retighten lock nut 2, lock roller 4 should then
be parallel to the edge of housing cover 5