Salter BPW-9154 MiBody Bluetooth Automatic Wrist Blood Pressure Monitor User Manual

Page 13

Advertising
background image

GB

13

Table 2 Guidance and manufacturer’s declaration – electromagnetic immunity – for all ME EQUIPMENT and ME SYSTEMS

Guidance and manufacture’s declaration – electromagnetic emission

The BPW-9154 is intended for use in the electromagnetic environment specified below. The customer or the user of the

BPW-9154 should assure that it is used in such an environment.

Immunity test

IEC 60601 test level

Compliance level

Electromagnetic

environment - guidance

Electrostatic

discharge (ESD)

IEC 61000-4-2

±6 kV contact

±8 kV air

±6 kV contact

±8 kV air

Floors should be wood, concrete or

ceramic tile. If floors are covered

with synthetic material, the relative

humidity should be at least 30%.

Electrical fast

transient/burst

IEC 61000-4-4

±2 kV for

power supply lines

N/A

Surge

IEC 61000-4-5

±1 kV line(s) to line(s)

±2 kV line(s) to earth

N/A

Voltage dips, short

interruptions and

voltage variations on

power supply input

lines

IEC 61000-4-11

<5% UT

(>95% dip in UT )

for 0.5 cycle

N/A

40% UT

(60% dip in UT )

for 5 cycles

N/A

70% UT

(30% dip in UT )

for 25 cycles

N/A

<5% UT

(>95% dip in UT)

for 5 sec

N/A

Power frequency 3A/m

(50Hz) magnetic

field

IEC 61000-4-8

3A/m

3A/m

Power frequency magnetic

fields should be at levels

characteristic of a typical location

in a typical commercial or hospital

environment.

NOTE UT is the a.c. mains voltage prior to application of the test level.