Page 13
GB
13
Table 2 Guidance and manufacturer’s declaration – electromagnetic immunity – for all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture’s declaration – electromagnetic emission
The BPW-9154 is intended for use in the electromagnetic environment specified below. The customer or the user of the
BPW-9154 should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for
power supply lines
N/A
Surge
IEC 61000-4-5
±1 kV line(s) to line(s)
±2 kV line(s) to earth
N/A
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
<5% UT
(>95% dip in UT )
for 0.5 cycle
N/A
40% UT
(60% dip in UT )
for 5 cycles
N/A
70% UT
(30% dip in UT )
for 25 cycles
N/A
<5% UT
(>95% dip in UT)
for 5 sec
N/A
Power frequency 3A/m
(50Hz) magnetic
field
IEC 61000-4-8
3A/m
3A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical location
in a typical commercial or hospital
environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.