Salter BPW-9154 MiBody Bluetooth Automatic Wrist Blood Pressure Monitor User Manual

Page 14

Advertising
background image

14

GB

Table 4 Guidance and manufacturer’s declaration – electromagnetic immunity – for ME EQUIPMENT and ME SYSTEMS that
are not LIFE-SUPPORTING

Guidance and manufacture’s declaration – electromagnetic emission

The BPW-9154 is intended for use in the electromagnetic environment specified below. The customer or user of the

BPW-9154 should assure that it’s used in such an environment.

Immunity test

IEC 60601

test level

Compliance

level

Electromagnetic

environment - guidance

Conducted RF

IEC 61000-4-6

Radiated RF

IEC 61000-4-3

3 Vrms

150 kHz to

80 MHz

3 V/m

80 MHz to

2.5 GHz

N/A

3 V/m

Portable and mobile RF communications equipment should

be used no closer to any part of the BPW-9154, including

cables, than the recommended separation distance

calculated from the equation applicable to the frequency of

the transmitter.
Recommended separation distance

d = 1.167 √

P

d = 1.167 √

P 80 MHz to 800 MHz

d = 2.333 √

P 800 MHz to 2.5 GHz

where

P is the maximum output power rating of the

transmitter in watts (W) according to the transmitter

manufacture and

d is the recommended separation

distance in metres (m).
Field strengths from fixed RF transmitters, as determined

by an electromagnetic site survey,

a

should be less than the

compliance level in each frequency range.

b

Interference may occur in the vicinity of

equipment marked with the following symbol:

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and

reflection from structures, objects and people.

a

Field strengths from fixed transmitters, such as base stations for radio (cellular / cordless) telephones and land

mobile radios, amateur radio, AM and FM radio broadcast cannot be predicted theoretically with accuracy. To assess

the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered.

If the measured field strength in the location in which the BPW-9154 is used exceeds the applicable RF compliance

level above, the BPW-9154 should be observed to verify normal operation. If abnormal performance is observed,

additional measures may be necessary, such as re-orienting or relocating the BPW-9154.

b

Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.

Advertising