Set-up and operating instructions – UVP EC3 Imaging System User Manual

Page 5

Advertising
background image

Gel Viewer

The viewer opens for safe viewing of samples. The safety glass al-
lows viewing without exposure to ultraviolet radiation.

Sample Platform and White Light Plate

The Sample Platform folds backwards against the back of the dark-
room cabinet. A White Light Plate is built into the platform and is
covered by a cover plate when the white light is not required.

Set-Up and Operating Instructions

NOTE: DO NOT ATTEMPT TO CONNECT ANY WIRING
WHILE THE EQUIPMENT IS CONNECTED TO ANY
POWER SUPPLY.

CAUTION!

Do not install the EC3 Imaging System in places with high moisture,
dust or high temperature.

Do not use any oil or petroleum based cleaner for the cabinet. Use
only mild soap or detergent solution for cleaning. Ensure that the
system is turned OFF during cleaning.

Set-Up

1.

Connect the camera and mount to the darkroom as specified in the
camera documentation manual.

2.

Connect the power cord to the main power port which is located on the
back-top of the darkroom cabinet.

3.

For shipping purposes, the chemi tray/white light platform is secured
to the back wall of the darkroom to prevent movement in shipping.

To release the platform, use a Philips head screwdriver to remove the
screws on the two sides of the platform.

3

EC3 Imaging System

Gel Viewer

Sample Platform, White Light
Plate and Transilluminator Tray

Screws