Troubleshooting – UVP EC3 Imaging System User Manual

Page 8

Advertising
background image

Troubleshooting

No power to darkroom cabinet.

1.

Recheck main power cord connections to the both the dark-
room and the wall power (surge protector).

2.

Check the fuse located on the main power port. If the darkroom
continues to blow fuses, call UVP Technical Support Depart-
ment.

Transilluminator will not turn on.

1.

Check to make sure the darkroom cabinet’s door is completely
closed. There is a UV exposure safety cut-off switch that turns
the transilluminator off when the darkroom cabinet’s door is
opened.

2.

Check the darkroom cabinet’s main power switch is lit. If not,
refer to “No power to darkroom cabinet” above.

3.

Be sure the transilluminator switch is on. When the switch on
the transilluminator is lit green, the unit is receiving power!

4.

Be sure the transilluminator control switch on the darkroom’s
front panel is selected to either “safety switch” or “always on.”

5.

Call UVP Technical Support Department with any technical ques-

tions.

6

EC3 Imaging System

Advertising
This manual is related to the following products: