Specifications


The FirstLight UV illuminator features a powder coat housing base and stainless steel frame. The
stainless steel frame is coated with an antimicrobial substance that resists bacterial growth. The
filter assembly is a unique combination of phosphored and filtered plates. The unit includes:

UV blocking cover

UV transmitting Gel-Tray with cm markings

Gel-Cutter tool

Fluorescent target

Physical dimensions of the FirstLight UV illuminator:

Width: 14.0” (356 mm)
Depth: 11.0” (279 mm)
Height: 5.63” (143 mm) Height includes UV blocking cover

Nominal

Model Part Number

Volt/Hz/Amp

Wavelength

Filter Size


FI-26X 95-0364-01

100-115/50-60/1.5

302 nm

25 x 26cm

FI-26X 95-0364-02

230/50/0.8

302 nm

25 x 26cm

FI-20

95-0365-01

100-115/50-60/1.5

302 nm

20 x 20cm

FI-20

95-0365-02

230/50/0.8

302 nm

20 x 20cm

FI-26

95-0366-01

100-115/50-60/1.5

302 nm

21 x 26cm

FI-26

95-0366-02

230/50/0.8

302 nm

21 x 26cm


Transilluminator Operation

Safety Precautions

Caution: The FirstLight UV illuminators are powerful sources of UV radiation that will cause
damage to unprotected eyes and skin. Before operating any unit, ensure all personnel in the area
are properly protected. It is preferable that the illuminator be installed and operated in a darkroom
where access and exposure is limited while the unit is in operation. If a darkroom is unavailable,
UVP offers imaging darkroom systems which provide protection from accidental exposure. For
information on the systems, contact UVP.

Unless configured with another system to create a darkroom environment, the illuminator is
equipped with a UV blocking cover. Even though this cover blocks the ultraviolet radiation emitted
by the unit, UV blocking eyewear should be worn as well. UVP has a complete line of UV blocking
eyewear: spectacles, goggles, and faceshields designed for this purpose. For part numbers, see
the Accessories section.

Caution: The grid lamp contains Mercury. Dispose as hazardous waste according to local, state,
and federal codes. MSDS sheet available upon request.