Wait Function ON/OFF: This function temporarily suspends counting time of the
program operation when influence of loads in the chamber or external disturbance such
as door operation has the chamber temperature failed in achieving the preset desired
temperature until the chamber temperature reaches the preset value. You may either
turn ON or OFF the function.