9special adjustment for measuring and adjusting co, Levels – Bosch 250 SXO NG User Manual

Page 27

Advertising
background image

6 720 607 440

Special adjustment for measuring and adjusting CO2 levels

27

9

Special adjustment for measuring and adjusting CO

2

levels

The CO

2

can only be adjusted by a certified gas

technician with a calibrated CO

2

analyzer.

Static Gas Pressure:

“ WC

P1 Operating Pressure:

“ WC

The P1 minimum operating pressure is 5" WC for
Natural Gas and 11"WC for Propane. Do not proceed
in adjusting CO

2

until pressure has been verified to be

at or above these levels, but not to exceed 14” WC.

A. Once Pressure is good

B

Turn ON/OFF switch to the OFF (O) position.

B

Remove plug by prying with a small flathead
screwdriver. Fig. 35

B

Insert CO

2

analyzer probe into the measuring port on

the front right of the vent cap at an angle. The tip of
the probe should be inserted to the middle of the vent
cap (approx 5" inserted).

Note: Vent cap must be in place in order to measure
CO

2

.

Fig. 35
B

While holding the Program (M) button, move the ON/
OFF switch to ON (I) position (see Fig. 36). As soon
as ‘188’ flashes on the display, release the Program
button. The display should now read P2.

Fig. 36

B. Measuring CO

2

(Cover Installed):

B

Open hot water taps to achieve a flow rate of at least
4 gallons per minute. (1 tub and 2 sinks should be
sufficient).

B

Record the CO

2

reading in P2 below. (Analyzer

reading may take several minutes to stabilize).

B

Press the

button until P1 appears. Unit will ramp

up to high fire and the flow should increase.

B

Record the CO

2

reading in P1 on the next page.

P2 CO

2

Reading:

% CO

2

P1 CO

2

Reading:

% CO

2

Compare your readings to those found in table 1 under
the “With Front Cover” column. If CO

2

readings are off

make adjustments as outlined below.
Note: The “Without Front Cover” column gives
approximate values with the cover off to ease the
adjustment process. Final readings should be taken
with the cover on.

C. Adjusting CO

2

:

Note: P1 adjustment will change the P2 reading.
Confirm the P1 value BEFORE adjusting the P2 level.

1. P1 CO

2

level off:

B

Loosen yellow painted Philips screw (1) and cover
should rotate down (2) revealing a brass slotted
screw. Fig. 37.

B

Adjusting the slotted screw counter clockwise will
raise P1 CO

2

levels and clockwise will lower P1

CO

2

levels. Adjustments to the slotted screw will

also change P2 CO

2

levels.

2. P2 CO

2

level off:

B

Remove yellow painted #40 Torx cover from the front
of the gas valve. (Fig. 38) A plastic #40 Torx screw
will be revealed.

B

Adjusting the plastic #40 Torx screw counter
clockwise will lower P2 CO

2

levels and clockwise

will raise P2 CO

2

levels.

Note: These screw adjustment are very sensitive and
may take several minutes to stabilize.

3. Verify both P1 and P2 are within the ranges specified
in table 1 under the “With Front Cover” column. Repeat
steps 1 and 2 as necesssary until CO

2

values are within

the specified ranges.

Caution: One factor that may affect
CO

2

levels is improper gas pressure.

Please see Chapter 2.9 for the
procedure to measure gas pressure
and record your findings below:

Advertising