Solving Problems…

If you see the letter “E” followed by a number, or if your
Meter beeps twice, an error has occurred. The following chart
gives you a list of possible errors and describes what they mean.

What You See

What it Means

What You Should Do

Low battery

Replace the batteries as soon as
possible.

E 1

Temperature out

Test only when the Meter tempera-

of range

ture is between 50°F and 104°F
(10°C and 40°C).

E 2

Test Strip underfill

Remove the Test Strip and repeat the
test with a new Test Strip.

E 3

Used Test Strip

Replace Test Strip with a new, unused
Test Strip. Wait until you see the
flashing blood drop before you add
blood sample.

E 4

Test Strip not

Remove the Test Strip and insert it

inserted correctly

properly. (See pg. 4.)

E 5 and above

All other errors

When you see the letter E followed
by a number that is not listed here,
remove Test Strip and retest with a
new Test Strip. If you continue to
experience difficulty, call our
Customer Service Department
(1-800-348-8100) for assistance.

The display

• Meter will turn off

• Insert new Test Strip to test, press

goes blank

after 3 minutes of

the M Button.

when the

inactivity. This fea-

meter is

ture prolongs your

turned on.

battery life OR

OR

• The battery may

• Replace the batteries if needed.

need to be replaced

LO

Test result is below

Repeat test. If LO appears again,

10 mg/dL (0.6 mmol/L)

contact your healthcare
professional (HCP) immediately.
CAUTION: Glucose levels below
50 mg/dL (2.8 mmol/L) may indicate a
potentially serious medical condition.