Opr. = operator; lay = lay user – Bayer HealthCare Ascensia Contour Blood Glucose Monitoring System User Manual

Page 39

Advertising
background image

35

Alternate

95% Confidence Interval

Site

Opr.

n

y =

Slope

/

Intercept

r

Palm

Lay

40

1.00x – 1.00

0.95 to 1.07 –11.43

to 4.26 0.986

Forearm

Lay

40

0.96x – 0.78

0.89 to 1.04 –13.15

to 12.93

0.973

Abdomen

Lay

39

0.85x + 11.32

0.76 to 0.95 –1.71

to 20.61

0.958

Thigh

Lay

40

0.97x – 4.10

0.88 to 1.07 –17.20

to 6.14

0.962

Opr. = Operator; Lay = Lay User

0

100

200

300

400

500

600

0

100

200

300

400

500

600

0

100

200

300

400

500

600

0

100

200

300

400

500

600

Ascensia

C

ONTOUR

Fingerstick Result (mg/dL)

Ascensia

C

ONTOUR

Fingerstick Result (mg/dL)

y=x

Ascensia

C

ONTOUR

Abdomen Result (mg/dL)

Ascensia

C

ONTOUR

Thigh Result (mg/dL)

y=x

C. Results from the Abdomen

D. Results from the Thigh

Passing and Bablok Regression Statistics / Pearson correlation