Altitude evaluation, Ontour – Bayer HealthCare Ascensia Contour Blood Glucose Monitoring System User Manual

Page 36

Advertising
background image

Altitude Evaluation

Performance was evaluated at an altitude of 10,000 ft.
Fifty-six samples were obtained from 54 different subjects
with diabetes. Three reagent lots were used with the
Ascensia C

ONTOUR

System to measure each subject’s capillary

blood glucose. After performing the Meter assays, additional
capillary blood was collected for a comparative blood glucose
(Yellow Springs Instruments Analyzer) and hematocrit
determination. The glucose range of the samples was 45 to
485 mg/dL, with an average concentration of 151 mg/dL. The
hematocrit range was 38 to 60%, with a mean of 48%.

32

Plasma Comparative Method Glucose Result (mg/dL)

Ascensia

C

ONTOUR

Result (mg/dL)

0

100

200

300

400

500

600

0

100

200

300

400

500

600

Lot 22B

Lot 22C

Lot 23B

y=x