Performance evaluations, Precision, For healthcare professionals) low control solution – Bayer HealthCare Ascensia Contour Blood Glucose Monitoring System User Manual

Page 27

Advertising
background image

Precision

1. Laboratory Precision: To assess the repeatability (within-

run precision) of the Ascensia

®

C

ONTOUR

Blood Glucose

Monitoring System, 100 blood glucose readings (10
readings on each of 10 instruments) were obtained with
venous blood specimens at five glucose concentrations. The
following table summarizes the mean and pooled within-
Meter %CV at each level.

2. Control Testing Precision: Precision results were obtained

from duplicate assays of Control Solutions at two diabetes
clinics with the Ascensia C

ONTOUR

System. Testing was

performed over 63 days. People with diabetes and
healthcare professionals (HCP) at each clinic evaluated two
reagent lots. One lot was evaluated at both clinics. The
tables below show the combined results for the lay users
and HCPs. The overall standard deviations (SD) and
coefficients of variation (CV) include the within run and
between run / day variations.

23

Duplicate

Overall Overall

Reagent

Assays

Mean

SD

CV

Lot

Operator

(n = )

(mg/dL)

(mg/dL)

(%)

22B

HCP

(n = 1)

35

42.2

1.33

3.2

Lay User

(n = 51)

76

42.4

1.83

4.3

22C

HCP

(n = 4)

39

43.1

1.14

2.6

Lay User

(n = 53)

78

43.0

1.67

3.9

23B

HCP

(n = 5)

74

42.6

1.14

2.7

Lay User

(n = 104)

154

41.8

2.02

4.8

Performance Evaluations

(for Healthcare Professionals)

Low Control Solution

Mean

46 mg/dL

92 mg/dL

144 mg/dL

235 mg/dL

376 mg/dL

%CV

2.4%

2.5%

2.8%

3.1%

2.4%