2 cleaning the spm, Figure 52, Spm – components – Metrohm 850 Professional IC Anion MCS Prep 3 User Manual
Page 103
■■■■■■■■■■■■■■■■■■■■■■
4 Operation and maintenance
850 Professional IC –Anion – MCS – Prep 3
■■■■■■■■
93
3 Connecting the SPM to the IC system
■
Reconnect the SPM to the IC system. If the capacity problems
remain, the SPM rotor A must be replaced (see Chapter 4.10.3.3,
page 95).
4.10.3.2
Cleaning the SPM
It may be necessary to clean the SPM in the following cases:
■
Increased backpressure on the connection tubings of the SPM.
■
Blockage of the SPM which cannot be eliminated (solutions can no lon-
ger be pumped through the SPM).
■
Jamming of the SPM which cannot be eliminated (SPM can no longer
be switched over).
1
2
3
4
5
Figure 52
SPM – Components
1
Union nut
2
SPM connecting piece (6.2835.010)
3
SPM rotor A (6.2835.000)
Cation exchanger.
4
SPM housing
5
Slot in the SPM housing
SPM cleaning
Clean the SPM as follows:
1 Disconnecting the SPM from the system
■
Switch off the instrument.
■
Disconnect capillary connections of the SPM from the system.
2 Dismounting SPM
■