6 installing the titration head, Installing the titration head, Figure 11 – Metrohm 814 USB Sample Processor User Manual

Page 28: Mounting the titration head

Advertising
background image

3.6 Installing the titration head

■■■■■■■■■■■■■■■■■■■■■■

20

■■■■■■■■

814 USB Sample Processor

Caution

Take care to ensure when mounting tubing and cables that there is no
traction on the drives while moving the lift or swiveling the robotic arm.
This could lead to overloading of and possible damage to the drive.

Remove the clips of the two lowest chain links when you install the
rinsing and aspiration tubing.

3.6

Installing the titration head

Mounting the titration head

6.1458.010

Figure 11

Mounting the titration head

1 Place the titration head (e.g. 6.1458.010) into the opening of titra-

tion head holder and screw tight with the enclosed screws.

Note the orientation of the titration head. The arrow on the titration
head marks the opening for the stirrer and has to point to the front,
see figure.

Installing accessories for the titration

The following figure shows the installation of the accessories in case rins-
ing and aspirating of sample vessels with the aid of integrated or external
pumps is used. Proceed as follows: