3 Installation

814 USB Sample Processor

■■■■■■■■

31

Switch the barcode reader to programming mode and make the following
settings:

1

Select the keyboard layout for the desired country (USA, Ger-
many, France, Spain, Switzerland (German)). This setting must
match the setting in the control software.

Make sure that the Ctrl characters (ASCII 00 to 31) are allowed to
be sent.

Adjust the settings so that the ASCII character 02 (STX or Ctrl B) is
sent as the first character as "Preamble" or "Prefix Code".

Adjust the settings so that the ASCII character 04 (EOT or Ctrl D) is
sent as the last character as "Postamble" or "Record Suffix" or
"Postfix Code".

Exit programming mode.

3.10

Mounting the drip pan

Serious damage to the instrument or a danger to the user can occur if
chemicals or liquid samples are spilled. The use of the drip pan
(6.2711.060) is recommended in order to avoid such incidents.

Mounting the drip pan

1

2

Figure 24

Installing the drip pan

Install the drip pan as follows:

1 Fasten the tubing enclosed to the drainage nipple on the drip pan

and then guide the free end of the tubing into a waste container.

2 Place the drip pan on the assembly rail of the turntable as shown in

the figure.