2 connecting a stirrer or titration stand, Connecting a stirrer or titration stand, Figure 20 – Metrohm 814 USB Sample Processor User Manual
Page 36: Connecting msb stirrer, Figure 21, Rod stirrer and titration stand
3.8 Connecting MSB devices
■■■■■■■■■■■■■■■■■■■■■■
28
■■■■■■■■
814 USB Sample Processor
3.8.2
Connecting a stirrer or titration stand
You can use a magnetic stirrer 801 Stirrer or 803 Ti Stand (stirring "from
below") or the 804 Ti Stand with a rod stirrer 802 Stirrer (stirring "from
above").
Connect a stirrer or a titration stand as follows:
1 Connect a stirrer or titration stand
■
Exit the control software.
■
Connect the connection cable of the magnetic stirrer or of the
titration stand to one of the sockets marked with MSB on the
rear of the control instrument.
■
If desired, connect the rod stirrer to the stirrer socket (with stirrer
symbol) of the titration stand.
■
Start the control software.
USB 1
Contr.
MSB 2
MSB 3
Figure 20
Connecting MSB stirrer
Figure 21
Rod stirrer and titration stand