3.11 Attaching the sample rack

■■■■■■■■■■■■■■■■■■■■■■

32

■■■■■■■■

814 USB Sample Processor

3.11

Attaching the sample rack

1

Figure 25

Attaching the rack

Put the rack into place as follows:

1 Carefully center the rack on the turntable. The guide bolts on the

turntable must engage with the openings in the bottom of the rack.
Tip: hold the rack in such a way that the printed Metrohm logo is
legible horizontally.

2 Carry out the [Rack Reset] function in the manual operation of the

control software.

The rack is moved into starting position. The magnet code of the
rack is read by the instrument during this process. The white arrow in
figure 25 indicates the position of the magnet holder. The six-digit
magnet code is used to identify the rack type. The sample positions
and any special positions on the rack are defined along with the rack
type.

3.12

Mounting the safety shield

Warning

It is imperative that the safety shield be installed before the first time
the 814 USB Sample Processor is used. The device is not permitted to
be operated without a safety shield.