3.6 Connecting sensors

■■■■■■■■■■■■■■■■■■■■■■

28

■■■■■■■■

907 Titrando

6 Insert the filled 6.1403.040 adsorber tube on the right of the elec-

trode into the remaining opening and then tighten the screw nipple
until it seals.

3.6

Connecting sensors

The measuring interface contains the following measuring inputs:

Ind. for a potentiometric electrode (pH, metal or ion-selective elec-
trode)

Ref. for a separate reference electrode

Temp. for a temperature sensor (Pt1000 or NTC)

Pol. for a polarizable electrode

iConnect for an iConnect (measuring interface for electrodes with
integrated data chip, so-called iTrodes)

3.6.1

Connecting a pH, metal or ion-selective electrode

Connect the pH, metal or ion-selective electrode as follows:

1 Plug the electrode plug into the Ind. socket of the 907 Titrando.

Figure 13

Connecting a pH, metal or ion-selective electrode

NOTE

The electrode cable is protected against accidental disconnection
of the cable by means of a pull-out protection. If you wish to pull
out the plug again, you will first need to pull back the outer plug
sleeve.