3 installation of the low pressure gradient, Installation of the low pressure gradient – Metrohm 850 Professional IC Anion MSM-HC MCS LP-Gradient User Manual

Page 26

Advertising
background image

3.3 Installation of the low pressure gradient

■■■■■■■■■■■■■■■■■■■■■■

16

■■■■■■■■

850 Professional IC–Anion – MSM-HC – MCS – LP Gradient

3.3

Installation of the low pressure gradient

The installation of the 850 Professional IC – Anion – MSM-HC – MCS – LP
Gradient must be adjusted slightly in order to be operated as a low pres-
sure gradient system. The following installation instructions list the neces-
sary work steps in brief. At each step you will find cross-references to
more detailed installation instructions for individual components, should
you require such aids.

Installing the low pressure gradient

1 Expanding the instrument to low pressure gradient

See Chapter 3.11, Page 38

Loosen the 6.1834.090 connection tubing from the output of the
eluent degasser.

Dismount the coupling (21-9) and pump head inlet capillary
(21-7) of the high pressure pump.

Use the short pressure screw to tightly screw the 6.2758.020 mix-
ing coil for low pressure gradients directly onto the input of the
high pressure pump, (see Chapter 3.12.2, page 43) and the lon-
ger pressure screw to screw it tightly to the output of the mixing
valve (see "Connecting mixing coil for low pressure gradients",
page 40)
.

Screw on the 6.1834.120 connection tubing to the output of the
eluent degasser (see "Connecting the eluent degasser", page
37)
and connect with input A of the mixing valve
(see "Connect-
ing connection tubings", page 41)
.

Equip two 6.1834.080 eluent aspiration tubings (see "Assembling
eluent aspiration tubing", page 33)
and screw tightly to the elu-
ent degasser inputs in the low pressure gradient (see "Connecting
the eluent degasser", page 37)
.

Advertising