15 sample degasser, Figure 25, Pulsation damper – connection – Metrohm 850 Professional IC Anion MSM-HC MCS LP-Gradient User Manual

Page 58

Advertising
background image

3.15 Sample degasser

■■■■■■■■■■■■■■■■■■■■■■

48

■■■■■■■■

850 Professional IC–Anion – MSM-HC – MCS – LP Gradient

3

4

5

2

3

6

1

2

Figure 25

Pulsation damper – Connection

1

Connection capillary
Connection to the inline filter.

2

Fastening screws

3

PEEK pressure screws, short
6.2744.070

4

Holder for pulsation damper

5

Pulsation damper 6.2620.150

6

Connection capillary
Connection to the injection valve.

3.15

Sample degasser

The sample degasser removes gas bubbles and disolved gases from the
sample. For degassing, the sample flows into a vacuum chamber through
a special fluoropolymer capillary.

Gas bubbles in the sample lead to poor reproducibility, as the quantity of
sample in the sample loop would not always be the same. Samples (con-
taining gas) should therefore be degassed before injection. For this the
sample is sucked through a degasser chamber before injection, whereby
any gas bubbles are automatically removed.

NOTE

When using the sample degasser, the rinsing time extends by at least 2
minutes.