Metrohm 940 Professional IC Vario ONE/ChS User Manual
Page 45
■■■■■■■■■■■■■■■■■■■■■■
3 Installation
940 Professional IC Vario ONE/ChS (2.940.1200)
■■■■■■■■
37
Connecting the rinsing solution inlet with STREAM
1 Connect the detector output capillary and the capillary labeled rins-
ing solution to each other using a coupling (6.2744.040) and two
pressure screws (6.2744.070).
Connecting the rinsing solution inlet to the Dosino
The rinsing solution capillary can be connected to the Dosino to which the
regenerant capillary is already connected.
For this step you need the following accessories:
■
Accessory kit: Flex/Vario: ChS (6.5000.030)
■
IC equipment: Dosino Regeneration (6.5330.190)
1 Fasten the capillary labeled rinsing solution to the outlet of an inline
filter (6.2821.120) using a pressure screw (6.2744.070).
2 Use a pressure screw (6.2744.070) to fasten one end of the PEEK
capillary (6.1803.030) to the inlet of the inline filter.
3 Use a pressure screw (6.2744.070) and an adapter (6.2744.080) to
fasten the other end of the PEEK capillary to Port 1 of the Dosino.