6 troubleshooting, 1 problems and their solutions, Problems and their solutions – Metrohm 850 Professional IC Cation - Gradient User Manual

Page 82

Advertising
background image

6.1 Problems and their solutions

■■■■■■■■■■■■■■■■■■■■■■

74

■■■■■■■■

850 Professional IC Cation – HP-Gradient

6 Troubleshooting

6.1

Problems and their solutions

Problem

Cause

Remedy

Marked drop in
pressure

Leak in the system.

Check all capillary connections and seal leaks,
if necessary (see Chapter 3.5, page 16).

Marked rise in pres-
sure

6.2821.120 inline filter
blocked.

Replace 6.2821.130 filter (see Chapter 5.6,
page 67)
.

Guard column – blocked.

Replace guard column (see Chapter 3.17,
page 47)
.

Separation column –
blocked.

Regenerate separation column (see Chap-
ter 5.11.4, page 72)
.

Replace separation column (see "Connect-
ing and rinsing the separation column",
page 49)
.

Note: Samples should always be microfiltered
(see Chapter 5.7, page 69).

Injection valve – valve
blocked.

Have the valve cleaned (by Metrohm service
technicians).

Drift of the baseline

Thermal equilibrium not yet
attained.

Condition instrument with the column thermo-
stat (see Chapter 3.15, Page 43) switched on .

Leak in the system.

Check all capillary connections and seal leaks,
if necessary (see Chapter 3.5, page 16).

Eluent – Evaporation of the
organic solvent in the elu-
ent.

Check the eluent bottle attachment (see Figure
13, page 29)
.

Peak areas lower
than expected

Sample – Leak in the sam-
ple path.

Check the sample path.

Sample – Blockage in the
sample path.

Check the sample path.

Sample – Sample loop not
(completely) filled.

Prolong the sample transfer time.

Sample – Gas bubbles in
the sample.

Use the sample degasser (see Chapter 3.13,
page 39)
(if available).