10 column thermostat, 11 mains connection, 12 interfaces – Metrohm 850 Professional IC Cation - Gradient User Manual

Page 88

Advertising
background image

7.10 Column thermostat

■■■■■■■■■■■■■■■■■■■■■■

80

■■■■■■■■

850 Professional IC Cation – HP-Gradient

7.10

Column thermostat

Type

Peltier technique thermostat for two intelligent separation columns

Adjustable tem-
perature range

0…+ 80 °C, in increments of 0.1 °C

Heating

Ambient temperature +50 °C

Cooling

Ambient temperature –20 °C

Temperature
reproducibility

± 0.2 °C

Stability

< 0.05 °C

Heating up time

< 30 minutes from 20 to 50 °C

Cooling time

< 40 minutes from 50 to 20 °C

7.11

Mains connection

Required voltage

100…240 V ± 10 % (auto-sensing)

Required fre-
quency

50…60 Hz ± 3 (auto-sensing)

Power consump-
tion

65 W for typical analysis application

25 W standby (detector to 40 °C)

Power supply unit

Up to 300 W maximum, electronically monitored

internal fuse 3.15 A

7.12

Interfaces

USB

Input

1 USB upstream, type B (for connection to the PC)

Output

2 USB downstream, type A

MSB

2 MSB Mini Din 8-pin (female) (for Dosino, stirrer, remote lines, ...)

Caution

When connecting an instrument to the MSB connector you must
switch off the 850 Professional IC .

Detector

2 DSUB-15-pin high density (female)

Column recogni-
tion

3 (including 2 in the column thermostat (see Chapter 3.15, page 43))