2.2 Setting up the KF titration cell

■■■■■■■■■■■■■■■■■■■■■■

6

■■■■■■■■

870 KF Titrino plus

2.2

Setting up the KF titration cell

Mounting the KF titration cell

6.2013.010

6.1414.030

6.1415.220

6.1903.020

1

2

3

Install the KF titration cell as follows:

1 Screw the clamping ring 6.2013.010 tightly to the support rod.

2 Mount the vessel lid 6.1414.030 of the KF titration cell to the sup-

port rod. Keep the locking lever pressed down until it can be released
at the desired position.

3 Fasten the titration vessel 6.1415.220 (or 6.1415.250) with a stirring

bar 6.1903.020 (or 6.1903.030) inside on the vessel lid. Fold back
the holding bracket upwards while doing so. The markings on the
vessel lid and on the plastic ring must be aligned above one another.
Afterwards, press the holding bracket downward in order to mount
the titration vessel. The levers of the holding bracket must enclose
the cams of the plastic ring on the titration vessel in order to ensure
a secure hold.