4 connecting the pump, Connecting the pump – Metrohm 870 KF Titrino plus Installation User Manual

Page 18

Advertising
background image

2.4 Connecting the pump

■■■■■■■■■■■■■■■■■■■■■■

12

■■■■■■■■

870 KF Titrino plus

3 Place the bottle top on the clear glass bottle 6.1608.030 (or on a dif-

ferent bottle with GL 45 thread) and screw it tightly.

4 Place the olive 6.1808.050 in the M8 connector of the bottle top and

screw it tightly.

5 Place the PVC tubing 6.1801.120 on the olive.

6 Insert the stopper 6.1446.090 into the remaining opening on the

bottle top.

7 Introduce the PTFE cannula 6.1819.050 into the opening of the stop-

per.

8 Insert a PTFE tubing 6.1805.200 with M8 connector into the opening

of the stopper and screw it tightly.

2.4

Connecting the pump

Connecting tubings to the pump

The pump connectors of the 803 Ti Stand are located on the rear side.

1

2

Please ensure that you connect the tubings to the pump correctly. Pro-
ceed as follows:

1 Fasten the PTFE tubing from the aspiration bottle (clear glass)

securely to the Waste connector (left-hand connector).

2 Fasten the PTFE tubing from the solvent bottle (amber glass with dry-

ing tube) securely to the Solvent connector (right-hand connector).