Application Bulletin 356

Installation instruction for «VoltIC pro I»

Page 11 of 18

Enter «stopp» in the field «End Print Job» under «Printer

Commands»

Click «Apply» and «OK».

3.2

Installation of MagIC Net

TM

All programs must be shut down first. Install MagIC Net

TM

(MagIC Net

TM

CD). All the standard directories proposed by

the program should be accepted. Restart windows.

Connected USB devices are automatically recognized when

MagIC Net

is started. The Microsoft Installation Wizard is

accepted and executed. After confirmation of the

automatically generated request, the devices are stored in

the configuration. The devices are predefined as «850

Professional IC 1» and «858 Professional Sample

Processor 1». Name them accordingly, if other names

appear in your configuration (e.g. due to changed setting on

your computer).

3.3

Configuration of MagIC Net

TM

In the window configuration, the 850 Professional IC, the

858 Professional Sample Processor and the columns

Metrosep A Supp 5 - 150/4.0 and Metrosep C 4 - 100/4.0

are visible. Add and define the eluents and the suppressor

solutions.

The settings for the rack require a «work position» fitting to

the respective rack (e.g. 125 mm for rack 6.2041.450).

Also define an «External 1» position to access the rinsing

station for rinsing the needle. The needle position is in the

inner compartment which corresponds to an angle of ~116°

and a work position of 120 mm. The exact angle depends

on your installation (see chapter 2.4).

For the outer compartment, an external position 2 can be

defined with an angle of ~110° and a work position of 60

mm. The external position 2 could be used as drain position

(e.g. to expel sample directly to the waste), however, it is

not carried out during the time program of method VoltIC pro

I.