Application Bulletin 356

Installation instruction for «VoltIC pro I»

Page 6 of 18

c) Establish the tubing connections at the measuring

head of the 797 VA Computrace:

2.4 858 Professional Sample Processor

For a detailed description, please refer to the 858

Professional Sample Processor manual. In general, you will

have to do the following: Plug in the Swing Head connection

cable, and plug in the power supply cable. The sample

processor initializes and lifts its Swing Head, once it is

connected to the IC device and the MSB connection is

recognized. Afterwards, it is possible to mount the retaining

plate (manual chapters 3.2 - 3.4).

Mount the Dosino holder (6.2057.040) on the side of the

tower and attach a 10 mL Dosing Unit (6.3032.210) and a

800 Dosino (2.800.0010) to it.

A rinsing station is mandatory for this setup.

The rinsing station is mounted onto the sample processor,

which is easiest when the rack is temporarily removed. The

exact position can be adjusted by directing the needle into

the rinsing station with the manual control (in the window

manual). The inner compartment approximately relates to a

swing head angle of ~116°, the outer one approximately

relates to ~110°. Memorize your angle settings.

Fig.: Rinsing station parts

A tubing is fixed at the bottom of the rinsing station to drain

the waste water. A bottle of ultrapure water is needed as

reservoir for the rinsing solution. The rinsing solution is

transported by one channel of the peristaltic pump of the

sample processor.

Mount the rinsing station on the sample processor.

Connect ultrapure water reservoir with the peristaltic pump

(via pump tubing (6.1826.360) and PEEK screws / tubing

olives) using a PTFE capillary (part of 6.1803.030). Then

connect the peristaltic pump (via pump tubing and PEEK

Screws / tubing olives) to the inner compartment of the

rinsing station also using a PTFE capillary (part of

6.1803.030).

The outer compartment of the rinsing station is connected to

the waste container.